Thermo Scientific Glacios 2

Overview:
The Glacios 2 cryo-TEM is optimized for high-throughput screening, single-particle data collection, and cryo-electron tomography (cryoET) of cryogenic samples.Features:
- Falcon 4i direct electron detector
- Ceta 16M CMOS camera
- SmartEPU with Quality Monitor, Multigrid, and CryoFlow
- Tomography 5
- 12-sample Autoloader
- Aberration-free image shift (AFIS)
- Fringe-free imaging (FFI)
Imaging Parameters:
- Accelerating voltage: 200kV
- Spherical aberration (Cs): 2.7mm
Normal Magnification Pixel Size (Å/pix) Benchmark Resolution (Å)+ 73,000x 1.92 — 92,000x 1.53 — 120,000x 1.166** 2.64 150,000x 0.94 — 190,000x 0.7159** 2.10 * 190,000 x magnification maximizes the field of view relative to the area exposed with the electron beam, covering the greatest area on the sample grid in the shortest time possible.
** Pixel sizes calibrated by comparison of a benchmark apoferritin reconstruction to a high-resolution x-ray crystal structure. Other pixel sizes are measured using a standard cross-grating replica grid.
+ Single-particle reconstruction of equine spleen apoferritin (Sigma Aldrich).
-
Thermo Scientific Talos F200C

Overview:
The Talos F200C is suited for room-temperature imaging of negatively stained samples; screening, single-particle data collection, or cryo-electron tomography (cryoET) of cryogenic samples; and microcrystal electron diffraction (microED).Features:
- Apollo direct electron detector
- Ceta-F CMOS camera
- EPU
- Tomography 5
- SerialEM
Imaging Parameters:
- Accelerating voltage: 200kV
- Spherical aberration (Cs): 2.7mm
Normal Magnification Ceta-F pixel size (Å/pix)* Apollo pixel size (Å/pix)* 13,500 — 2.284 17,500 — 1.808 22,000 6.7 1.415 28,000 5.2 1.116 36,000 4.1 0.8932 45,000 3.3 0.7038 57,000 2.6 0.5594 73,000 2.0 0.4394 92,000 1.6 0.349 120,000 1.3 — 150,000 0.99 — 190,000 0.77 — * Pixel sizes are physical resolution (4k x 4k) for the Ceta-F and super-resolution (8k x 8k) for the Apollo by default.
-
Plunge-Freezing Systems
Overview:
The Vitrobot automates the process of blotting and plunge-freezing samples for cryo-TEM imaging to improve consistency and reproducibility. Typical specimens include proteins, viruses, vesicles, bacteria, and cells.
Features:
- Humidity and temperature control
- Double-sided blotting
Gatan Cryoplunge
Features:
- Single- or double-sided blotting
-
PIE Scientific Tergeo EM Plasma Cleaner
Features:
- 20 customizable template recipes
- Inlets for three process gases
- Water vapor attachment for steady supply of hydrophilic functional groups
- Remote plasma generation for gentler cleaning/modification of TEM grids
-
Pelco EasiGlow Glow Discharge System
Features:
- Customizable glow discharging process
- Single-button operation
-
Side-Entry Cryogenic Sample Holders
Gatan Model 626

Features:
- ~4hr liquid nitrogen holding time
- ≥3.4Å resolution
Gatan “Elsa” Model 698

Features:
- Ultra-low profile for high tilt angles
- ~8hr liquid nitrogen holding time
- ≥2.3Å resolution
Simple Origin Model 201

Features:
- Dual grid holder: one standard TEM grid and one Autogrid
- ~8-10hr liquid nitrogen holding time
- ≥2.4Å resolution
-
Computer Hardware
GPU server (Jupiter)
Overview:
Jupiter is a multi-GPU computing server acting as the main cryoSPARC worker node used for on-the-fly data processing, which matches the pace of data collection on the Glacios 2.
Features:
- 4 x NVIDIA GeForce RTX 4060 Ti 16GB GPUs
- 32-core 64-thread AMD EPYC 7543P Milan CPU
- 256 GB SuperMicro DDR4 RAM
GPU workstation (Saturn)
Overview:
Saturn utilizes a single cutting-edge GPU for tilt-series processing and tomographic reconstructions as well as single-particle data processing.
Features:
- NVIDIA RTX PRO 6000 Blackwell Max-Q Workstation Edition 96GB GPU
- 32-core 64-thread AMD Ryzen Threadripper Pro 9975WX CPU
- 256 GB Kingston DDR5-6400 MHz ECC Reg. RAM